Reproducibility of semi-quantitative assessment of aortic valve calcification and valve motion on echocardiography: a small-scale study

Abstract

Background

Aortic stenosis (AS) is the most common degenerative valve disease in high income countries. While hemodynamic metrics are commonly used to assess severity of stenosis, they are impacted by loading conditions and stroke volume and are often discordant. Anatomic valve assessments such as aortic valve calcification (AVC) and valve motion (VM) during transthoracic echocardiography (TTE) can offer clues to disease severity. The reliability of these semi-quantitatively assessed anatomic imaging parameters is unknown.

Methods

This is a retrospective study of semi-quantitative assessment of AVC and valve VM on TTE. TTEs representing a range of AS severities were identified. The degree of calcification of the aortic valve and the degree of restricted VM were assessed in standard fashion. AVC scores and valve motion were assessed by readers with varied training levels blinded to the severity of AS. Correlation and inter-reader reliability between readers were assessed.

Results

420 assessments (210 each for AVC and VM) were collected for 35 TTEs. Correlation of AVC for imaging trainees (fellows and students, respectively), ranged from 0.49 (95% CI 0.18–0.70) to 0.62 (95% CI 0.36–0.79) and 0.58 (95% CI 0.30–0.76) to 0.54 (95% CI 0.25–0.74) for VM. Correlation of anatomic assessments between echocardiographer-assigned AVC grades was r = 0.76 (95% CI 0.57–0.87)). The correlation between echocardiographer-assigned assessment of VM was r = 0.73 (95% CI 0.53–0.86), p < 0.00001 for both. For echocardiographer AVC assessment, weighted kappa was 0.52 (0.32–0.72), valve motion weighted kappa was 0.60 (0.42–0.78).

Conclusion

There was good inter-reader correlation between TTE-based semi-quantitative assessment of AVC and VM when assessed by board certified echocardiographers. There was modest inter-reader reliability of semi-quantitative assessments of AVC and VM between board certified echocardiographers. Inter-reader correlation and reliability between imaging trainees was lower. More reliable methods to assess TTE based anatomic assessments are needed in order to accurately track disease progression.

Clinical Trial Number

STUDY00003100.

Introduction

Aortic stenosis (AS) is the most common valve disease in those aged 75 and older and progresses through fibrotic and calcific disease stages that ultimately lead to restricted valve motion (VM), obstruction of blood flow, remodeling of the left ventricle, heart failure, and death if left untreated [1]. While valve hemodynamics are the primary imaging features on transthoracic echocardiogram (TTE) used to assess valve disease severity, there is increasing interest in anatomic assessments of valve dysfunction [2]. Valve calcification and restricted VM observed during TTE are hallmark features of this disease and adjunctive imaging features that can be used to help stage disease severity. These features have also been proposed as tools to aid with risk stratification and prognosis [3, 4]. The reliability and reproducibility of semi-quantitative assessments of valve calcification and valve motion on TTE are unknown [2, 5].

Aortic valve calcification (AVC) and restricted VM are features of degenerative AS that can be seen on TTE and ultimately contribute to progressive pressure overload on the left ventricle. AVC progresses at variable rates and has demonstrated strong associations with traditional cardiovascular risk factors [5, 6]. This calcification coupled with pro-fibrotic processes ultimately lead to fusion of valve leaflets and restricted VM [7]. These phases demonstrate variable progression with women showing lower rates of calcification for any given stage of disease [8]. Semi-quantitative assessments of aortic valve anatomy are routinely reported and have been used in studies of AS disease progression [2, 9, 10]. Regardless of the quantification technique used, the baseline load of calcium on the aortic valve may be the most well-validated marker regarding the severity and progression of AS [2].

AVC as assessed by computerized tomography (CT) is the most accurate, reproducible, and well-validated marker of AS severity (and disease progression) at later disease stages, and is helpful in the diagnostic evaluation of patients with discordant echocardiographic markers of AS [2]. The reliability of semi-quantitative AVC and VM assessments during TTE imaging is unknown and remains a critical question since these assessments are incorporated into disease staging guidelines [11]. Highly reliable and reproducible TTE imaging would be preferable to CT due to ease of access, low cost of imaging, and lack of ionizing radiation. Here we conduct a study to assess the reproducibility and inter-reader reliability of standard semi-quantitative assessments of TTE-based AVC and VM assessments across the spectrum of AS between individuals of different levels of training.

Materials and methods

TTE selection protocol

This was a retrospective single center study of TTEs representing the full spectrum of AS. The study population was identified using a Phillips® ISCV search tool. This EMR-based search tool was used to identify patients with AS. Imaging from patients with ‘none, mild, moderate, or severe’ AS as classified by a board-certified echocardiographer (cardiologist with additional fellowship in cardiac imaging) were obtained and de-identified and randomly selected [11]. Patients were excluded if image quality was poor, either due to patient characteristics or quality of the study. For this study, each included study was independently reviewed by a board-certified echocardiographer to confirm severity of AS and to confirm that the imaging views were evaluable. Patients with prosthetic valves and inadequate aortic valve visualization were excluded.

Image selection protocol

For each study, de-identified videos representing zoomed parasternal long axis (PLAX) and parasternal short axis at the level of the aortic valve (PSAX AoV) were used. During the course of this study these imaging views were displayed side by side for each reader. Imaging was displayed in a random order and readers were blinded to the reference AS severity label.

Image interpretation protocol

Images were individually reviewed by 2 cardiologists boarded in echocardiography, 2 cardiology fellows, and 2 medical students. The cardiology fellows are post-graduate years 4–6. The medical students receive basic ultrasound training in their pre-clinical education and received an additional didactic session regarding PLAX and PSAX views of the aortic valve with various stages of AV disease. Readers were asked to grade AVC in standard fashion using a scale from 1 to 4 (1, no calcification; 2, mild was defined as few areas of dense echogenicity with little acoustic shadowing; 3, moderate as multiple larger areas of dense echogenicity; and 4, severe as extensive thickening and increased echogenicity with a prominent acoustic shadow). Degree of VM was graded from 1 to 4 (1, normal motion; 2, mildly restricted motion; 3, moderately restricted motion; and 4, heavily restricted motion).

Statistical analysis

Reproducibility and inter-reader reliability were assessed. Reproducibility is defined as variation of the same measurement made on the same subject by different readers, while inter-reader reliability is defined as the ability for different readers to come to similar conclusions when shown the same image [10]. The correlation for AVC and VM grades assigned by readers with the same level of training was assessed using Spearman correlation with 95% confidence intervals. Inter-reader agreement was assessed using the Kappa statistic to compare agreement between two readers of the same training level. P-values of < 0.05 were considered statistically significant. Statistical analysis was performed using SAS 9.4 statistical software. Kappa ranges correspond to: ≤ 0 as indicating no agreement and 0.01–0.20 as none to slight, 0.21–0.40 as fair, 0.41– 0.60 as moderate, 0.61–0.80 as substantial, and 0.81–1.00 as almost perfect agreement [10].

Results

This study included imaging from 35 patients represented by 70 videos (PLAX and PSAX of AoV), 2 for each patient. 6 readers analyzed AVC and valve motion from 35 TTEs (210 AVC labels and 210 valve motion labels). The imaging cohort included patients with no AS (n = 5), patients with mild AS (n = 10), patients with moderate AS (n = 10), and severe AS (n = 10). The echocardiographic characteristics of the patient images are shown in Supplemental Table 1. The median age was 73 years (IQR 13). 71% of the patients were men. 88% of the study population was white.

AVC

Correlation between the echocardiographer-assigned AVC grades was r = 0.76 (95% CI 0.57–0.87 [p < 0.0001]) (supplemental Fig. 1). For fellows, correlation coefficients was r = 0.49 (95% CI 0.18–0.70 [p = 0.0027]) (supplemental Fig. 3). Correlation for medical students was r = 0.62 (95% CI 0.36–0.79 [p < 0.0001]) (supplemental Fig. 5). For inter-reader reliability between board certified echocardiologists, AVC weighted kappa = 0.52 (0.32–0.72). Inter-reader reliability for trainees ranged from 0.37 (0.14–0.61) to 0.5 (0.29–0.72).

Valve motion

Correlation coefficients for echocardiographer-assigned assessment of VM was r = 0.73 (95% CI 0.53–0.86) [p < 0.00001]) (supplemental Fig. 2). For fellows, correlation coefficients were r = 0.58 (95% CI 0.30–0.76 [p = 0.0002]) (supplemental Fig. 4). The correlation between medical students were r = 0.54 (95% CI 0.25–0.74 [p = 0.0007]) (supplemental Fig. 6). For inter-reader reliability between board certified echocardiologists, VM weighted kappa = 0.60 (0.42–0.78). Inter-reader reliability for trainees ranged from was 0.29 (0.11–0.48) to 0.49 (0.27–0.72) for VM.

Discussion

The main finding from this study is that there is good correlation for TTE-based assessments of AVC and VM though inter-reader reliability is modest. While anatomic assessments of valve morphology can be helpful in assigning AS grade and correlate with subsequent outcomes, standard TTE-based semi-quantitative assessments lack the precision and reproducibility needed to reliably track disease progression. More accurate assessment of AVC could provide better tools for assessing AS severity. These findings suggest the need for more accurate and reproducible TTE-based methods to assess aortic valve morphology.

TTE remains the primary imaging modality used to assess AS severity and its ability to assess both anatomic and hemodynamic changes associated with worsening AS makes it ideally suited for tracking disease progression [4, 11]. Contemporary clinical assessment of AS severity relies heavily on an integrative approach that combines a number of (mostly hemodynamic) imaging parameters in order to assign a summary severity grade. While there has been substantial interest in refining hemodynamic parameterization of this condition, these metrics are often discordant [9]. There has been less attention paid to the visual assessment of AVC or VM on TTE and a semi-quantitative approach to grading these metrics is commonly used in practice and has been integrated into imaging guidelines [11]. The data presented here suggest that the current approach may not yield reliable results.

There are some early efforts underway to improve the precision and reproducibility of TTE-based anatomic assessments of the aortic valve. In a previous single center study, a novel global calcium (GC) score, defined as the summed means of grayscale in 3 regions of interest (ROI) in PLAX and 5 ROI in PSAX was found to correlate with CT Agatston score [12]. Unfortunately, this analysis included only 14 patients with AS and to our knowledge has not been repeated. Investigators found similar success and application to echocardiographic measurement of GC in comparison to CT score [13, 14]. An investigation utilizing software that analyzes relative pixel brightness with the anechoic nature of blood as a control has been suggested as a potential tool for echocardiography experts to analyze AVC, finding a strong correlation between human expert and computer software assessment of calcium area for the same images [15].

More recently, a 2D-AVC ratio was defined as the average pixel density of the AV divided by the average pixel density of the aortic annulus was reported to correlate with hemodynamic severity of AS [16]. These results lay the groundwork for more accurate and reproducible assessments of AVC though remain limited in that they only use a diastolic PSAX image and do not integrate assessments of VM.

Restricted VM is an important anatomic feature of AS that has not yet been established as a prognostic variable in AS progression. As information on the sexual dimorphism of this condition emerges, and with evidence that calcification progresses at different rates (and to different thresholds) for men and women, a reliable and accurate method to assess VM could offer important insights about valve severity for those with fibrosis-dominant phenotypes. Additionally, as the age of artificial intelligence for echocardiography advances forward, attention to these morphologic features might improve confidence and agreement between providers when assessing severity of AS.

Semi-quantitative assessments in echocardiography are commonly used because quantitative analyses are time consuming and often discordant. As TTE workflows improve, higher reproducibility and accuracy should be expected. While automation tools continue to advance, they have yet to be trained on traditionally semi-quantitative measures. Adding these types of morphology assessments to the portfolio of automated measurements might ultimately help with more accurate and reproducible grading of AS, as well as other valve lesions [17]. This study included good quality TTE imaging and therefore represents a ‘best cases scenario’ for inter-reader correlation and reliability. In clinical practice, many factors impact the quality of TTE imaging, potentially worsening the real-world reliability of these parameters.

Conclusions

There was good inter-reader correlation for semi-quantitative assessment of AVC and valve motion however inter-reader reliability was modest. More reliable methods to assess TTE based anatomic assessments are needed in order to accurately track disease progression.

Data availability

Data is not publicly available to protect patient privacy. Data is stored on encrypted hard drive on Tufts Medical Center servers.

Abbreviations

AoV:

Aortic Valve

AS:

Aortic Stenosis

AVC:

Aortic Valve Calcification

CT:

Computerized tomography

GC:

Global Calcium

PLAX:

Parasternal Long Axis

PSAX:

Parasternal Short Axis

ROI:

Region of interest

TTE:

Transthoracic Echo

VM:

Valve Motion

References

  1. Lindman BR, Sukul D, Dweck MR, et al. Evaluating Medical Therapy for Calcific aortic stenosis: JACC state-of-the-art review. J Am Coll Cardiol. 2021;78(23). https://doi.org/10.1016/j.jacc.2021.09.1367.

  2. Pawade T, Sheth T, Guzzetti E, Dweck MR, Clavel MA. Why and how to measure aortic valve calcification in patients with aortic stenosis. JACC Cardiovasc Imaging. 2019;12(9). https://doi.org/10.1016/j.jcmg.2019.01.045.

  3. Rosenhek R, Klaar U, Schemper M, et al. Mild and moderate aortic stenosis: natural history and risk stratification by echocardiography. Eur Heart J. 2004;25(3). https://doi.org/10.1016/j.ehj.2003.12.002.

  4. Rosenhek R, Binder T, Porenta G, et al. Predictors of outcome in severe, asymptomatic aortic stenosis. N Engl J Med. 2000;343(9). https://doi.org/10.1056/nejm200008313430903.

  5. Farmer JA. Intensive lipid lowering with simvastatin and ezetimibe in aortic stenosis (the SEAS trial). Curr Atheroscler Rep. 2009;11(2):82–3.

    PubMed  Google Scholar 

  6. Chan KL, Teo K, Dumesnil JG, Ni A, Tam J. Effect of lipid lowering with Rosuvastatin on Progression of aortic stenosis. Circulation. 2010;121(2). https://doi.org/10.1161/circulationaha.109.900027.

  7. Messika-Zeitoun D, Bielak LF, Peyser PA, et al. Aortic valve calcification: determinants and progression in the population. Arterioscler Thromb Vasc Biol. 2007;27(3). https://doi.org/10.1161/01.ATV.0000255952.47980.c2.

  8. Clavel MA, Pibarot P, Messika-Zeitoun D, et al. Impact of aortic valve calcification, as measured by MDCT, on survival in patients with aortic stenosis: results of an international registry study. J Am Coll Cardiol. 2014;64(12). https://doi.org/10.1016/j.jacc.2014.05.066.

  9. Minners J, Allgeier M, Gohlke-Baerwolf C, Kienzle RP, Neumann FJ, Jander N. Inconsistencies of echocardiographic criteria for the grading of aortic valve stenosis. Eur Heart J. 2008;29(8). https://doi.org/10.1093/eurheartj/ehm543.

  10. Bunting KV, Steeds RP, Slater LT, Rogers JK, Gkoutos GV, Kotecha D. A practical guide to assess the reproducibility of echocardiographic measurements. J Am Soc Echocardiogr. 2019;32(12). https://doi.org/10.1016/j.echo.2019.08.015.

  11. Baumgartner H, Hung J, Bermejo J, et al. Recommendations on the Echocardiographic Assessment of Aortic Valve Stenosis: a focused update from the European Association of Cardiovascular Imaging and the American Society of Echocardiography. J Am Soc Echocardiogr. 2017;30(4). https://doi.org/10.1016/j.echo.2017.02.009.

  12. Gillis K, Bala G, Roosens B, et al. Clinical validation of an ultrasound quantification score for aortic valve calcifications. Int J Cardiol. 2018;252. https://doi.org/10.1016/j.ijcard.2017.07.020.

  13. Gaibazzi N, Baldari C, Faggiano P, et al. Cardiac calcium score on 2D echo: correlations with cardiac and coronary calcium at multi-detector computed tomography. Cardiovasc Ultrasound. 2014;12(1). https://doi.org/10.1186/1476-7120-12-43.

  14. Paulsen NH, Carlsen BB, Dahl JS, et al. Association between aortic valve calcification measured on non-contrast computed tomography and aortic valve stenosis in the general population. J Cardiovasc Comput Tomogr. 2016;10(4). https://doi.org/10.1016/j.jcct.2016.05.001.

  15. Elvas LB, Almeida AG, Rosario L, Dias MS, Ferreira JC. Calcium identification and scoring based on echocardiography. An exploratory study on aortic valve stenosis. J Pers Med. 2021;11(7). https://doi.org/10.3390/jpm11070598.

  16. Desai K, Slostad B, Twing A, Krishna H, Frazin L, Kansal M. Application of a Novel two-dimensional echocardiographic calcium quantification method to assess all severities of aortic stenosis. J Am Soc Echocardiogr. 2023;36(1). https://doi.org/10.1016/j.echo.2022.10.023.

  17. Krishna H, Desai K, Slostad B, et al. Fully automated Artificial Intelligence Assessment of aortic stenosis by Echocardiography. J Am Soc Echocardiogr. 2023;36(7):769–77. https://doi.org/10.1016/j.echo.2023.03.008.

    Article  PubMed  Google Scholar 

Download references

Acknowledgements

The authors are thankful to Tufts Medical Center for their support in data acquisition. No funding was received for this study. The data from this study is not publicly available. Ethical approval was received from Tufts IRB.

Funding

No funding was acquired for this study.

Author information

Authors and Affiliations

Authors

Contributions

D. B. collected data, wrote the manuscript, created supplemental Table 1, created supplemental Fig. 7, and is the corresponding author. B. K. conducted the statistical analysis and created supplemental Figs. 1–6. S. M. is one of the cardiology fellows, conducted data collection, and edited the manuscript. Y. D. is one of the cardiology fellows, conducted data collection, and edited the manuscript. S.H. M. was one of the medical students, conducted data collection, and edited the manuscript. J. F. was one of the medical students, conducted data collection, and edited the manuscript. J. L. was the sonographer who collected the TTEs, conducted data collection, created the powerpoint for data collection, and edited the manuscript. A. K. was one of the board certified cardiologists, aided in data collection, and edited the manuscript. B.C. D. was one of the board certified cardiologists, aided in data collection, and edited the manuscript. B. W. was the sponsor of the manuscript, and wrote and edited the manuscript.

Corresponding author

Correspondence to D. Balian.

Ethics declarations

Ethics approval and consent to participate

This study was ethically approved by the Tufts IRB.

Consent for publication

Not applicable.

Competing interests

The authors declare no competing interests.

Additional information

Publisher’s Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Electronic supplementary material

Below is the link to the electronic supplementary material.

Supplementary Material 1

Supplementary Material 2

Rights and permissions

Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article’s Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article’s Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/. The Creative Commons Public Domain Dedication waiver (http://creativecommons.org/publicdomain/zero/1.0/) applies to the data made available in this article, unless otherwise stated in a credit line to the data.

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Balian, D., Koethe, B., Mohanty, S. et al. Reproducibility of semi-quantitative assessment of aortic valve calcification and valve motion on echocardiography: a small-scale study. Echo Res Pract 11, 15 (2024). https://doi.org/10.1186/s44156-024-00050-3

Download citation

  • Received:

  • Accepted:

  • Published:

  • DOI: https://doi.org/10.1186/s44156-024-00050-3

Keywords